Datex-Ohmeda S/5 Anesthesia Delivery Units
FDA Recall Z-0937-03 · Posted 2003-08-28
Advertisement
Recalling FirmDatex-Ohmeda Inc
Recall StatusTerminated
Product CodeBSZ
Date Initiated2003-07-15
Date Terminated2005-11-08
Product Quantity544 units
Reason for Recall
Possible disruption of mechanical ventilation or disruption of anesthetic agent delivery.
Firm Action
Recall letters dated July 15, 2003 stated that a firm representative would contact the consignees to schedule a visit to replace CPU boards to correct the problem.
Root Cause
Other
Distribution Pattern
The product was distributed nationwide to hospitals in the Unites States.