MilCAM Recon III Model Family of IR Imagers that incorporate a Class I infared laser for Laser Range Finding (LRF).

FDA Recall Z-0937-2009 · Posted 2009-09-25

Recalling FirmFlir Systems, Inc.
Recall StatusTerminated
Product CodeREQ
Date Initiated2009-01-30
Date Terminated2015-06-02
Product Quantity120

Reason for Recall

LRF models were sold to non-DoD customers prior to certification and reporting as required in Title 21 of the Code of Federal Regulations, Subchapter J, Parts 1000 - 1050 (21 CFR 1000 - 1050). A filter defect allows transmission of laser energy such that the affected LRF would be Class 3B laser product, however the product is reported to CDRH as a Class I product.

Firm Action

Customers will be notified by a letter instructing them to stop using the LRF, contact a Customer Service Representative, and to detach and return the Laser Range Finder. For each returned LRF, FLIR will provide a certified Class I LRF.