Terminated

Hemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used in performing a quantitative, one -stage APTT. The citrate APTT test is used for evaluation of low doses of heparin anticoagulant (up to 1.5 units/mL), depending upon individual patient heparin sensitivity.

FDA Recall Z-0937-2014 · Posted 2014-02-06

Recalling FirmITC-Nexus Dx
Recall StatusTerminated
Product CodeGFO
Date Initiated2013-12-19
Date Terminated2014-10-30
Product Quantity25,965 curvettes (7,425 cuvettes US; 18,540 cuvettes OUS)

Reason for Recall

ITC has determined that some Citrate APTT cuvettes may recover higher than expected results in normal individuals.

Firm Action

The firm, ITC The Point of Care, sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 19, 2013 a long with an attached Health Professional Account Tracking Form via UPS to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to forward this communication to all those within your organization who need to be aware of this matter including any facility that you forwarded any of the affected product to; check all inventory to determine if you have any of the affected lots listed; if you have any affected lots, stop using them, remove from inventory and return to ITC; complete the attached Health Professional Account Tracking Form and return it within 10 days via fax to: 866-429-3132, email: [email protected] or mail to: ITC Technical Support, 20 Corporate Place South - Piscataway, NJ 08854 . It is important that you complete the form whether or not you have remaining inventory of these lots. In addition, contact ITC Technical Support immediately if you become aware of an unreported adverse event related to the use of affected J103C cuvettes. ITC will review your response form and will contact you to arrange for any inventory to be returned to ITC and ITC will replace all unused, returned product. If you have questions, please contact ITC Technical Support: Telephone 800-579-2255 or 732-548-5700, Extension 4011 or E-mail [email protected].

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Austria, Belgium, Canada, Serbia, Germany, Estonia, Italy, Spain and Saudia Arabia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →