Liquid Cardiac Control CQ5053

FDA Recall Z-0937-2018 · Posted None

Recall StatusTerminated
Product CodeJJY
Date Initiated2017-08-08
Date Terminated2018-09-05
Product Quantity450 kits

Reason for Recall

Randox has confirmed that Troponin T and Troponin I in Liquid Cardiac quality control CQ5053 lot 4069CK does not meet the shelf-life claim in the product labelling. As the control may now fall outside the provided control ranges the product is to be recalled from the field.

Firm Action

A recall notice was issued to customers on August 1, 2017 instructing them to discontinue use of product, discard remaining product for reimbursement and provide documentation of destruction.

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

Nationally