Stryker Orthobiologics Imbibe Bone Marrow Aspiration Needle, Model Numbers: 2090-0027 Needle (bullet-tip), 11 gauge x 4 inch 2090-0028 Needle (bullet-tip), 11 gauge x 6 inch 2090-0029 Needle (bullet-tip), 8 gauge x 6 inch 2090-0047 Needle (bullet-tip), 8 gauge x 8 inch 2090-0030 Needle (bullet-tip), Fenestrated, 8 gauge x 6 inch intended for use to aspirate bone marrow or autologous blood.

FDA Recall Z-0938-2015 · Posted 2015-01-07

Recall StatusTerminated
Product CodeKNW
Date Initiated2014-12-04
Date Terminated2015-07-28
Product Quantity89770

Reason for Recall

There is the potential for a breach in the inner or outer packaging pouches of all lots of the Imbibe Bone Marrow Aspiration Needle.

Firm Action

The firm, Stryker, sent an "URGENT PRODUCT RECALL" letter dated December 4, 2014, to end users/customers. The letter identified the product, problem, and actions to be taken. Customers were instructed to complete and return the URGENT MEDICAL DEVICE RECALL NOTIFICATION ACKNOWLEDGMENT FORM within 5 days via fax to 215-253-5020, contact other hospitals in your territory that have the affected products to arrange return of the product, and return the affected product to: Stryker Orthobiologics, 67 Great Valley Parkway, Malvern, PA 19355. If you have any questions, feel free to contact the Associate Manager, Quality Assurance and Regulatory Compliance at 610-640-1775 x 5299.

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution: US (nationwide) and country of: Canada.