ALung Technologies Hemolung Respiratory Assist System (RAS) with CME BodyGuard 323 Infusion Pump(EUA) - Product Usage: an off-the-shelf infusion pump and administration set to provide a 30 mL/hr continuous saline flush to the integrated blood pump bearing.
FDA Recall Z-0938-2021 · Posted None
Advertisement
Recalling FirmAlung Technologies Inc
Recall StatusTerminated
Product CodeQLN
Date Initiated2020-12-18
Date Terminated2021-05-17
Product Quantity101 units
Reason for Recall
CME America BodyGuard 323 Infusion Pumps (recalled by manufacturer) is a component to the Hemolung for EUA, and the infusion pump may have a delivery inaccuracy
Firm Action
ALung issued email notification "Recalled CME BodyGuard 323 Infusion Pump to Sapphire Pump Exchange" on 12/18/20 . Letter notifies users of the CME/Sapphire Infusion Pump exchange. Provided with this email are the instructions that will come with the infusion pump package. We wanted to give you a preview of what is coming prior to the packages arriving. Once received, we ask that you promptly remove the current CME pump, package and ship back to ALung.
These are needed to be returned to the manufacturer by the end of the year.
Reminder you may need to contact biomed for the electrical safety testing of the new Sapphire pump! Please follow your internal protocols.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.