Recalling FirmMisys Healthcare Systems
Recall StatusTerminated
Product CodeJQP
Date Initiated2005-05-10
Date Terminated2005-08-30
Product Quantity21
Reason for Recall
When a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more suceptibilities on the culture will be deleted.
Firm Action
Firm sent Product Safety Notice PSN 05-02 on May 10, 2005 which also provides a procedural work-around and an invitation to obtain a service pack to correct the software flaw.
Root Cause
Other
Distribution Pattern
Nationwide and United Kingdom and Denmark