Terminated

Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South McCaslin Blvd., Suite 100, Louisville, CO 80027

FDA Recall Z-0939-2011 · Posted 2011-01-25

Recall StatusTerminated
Product CodeGEI
Date Initiated2010-12-21
Date Terminated2011-02-08
Product Quantity27 units

Reason for Recall

Surgical equipment, activated by a wireless foot switch, may remain powered on when switch is no longer depressed.

Firm Action

Customers were notified by e-mail with an Advisory Notice on 12/21/2010. The Notice identified the affected product and described the issue. The Notice also explained what the potential hazard involved along with a mechanism to eliminate the potential hazard. Sound Surgical would arrange a service call to implement the correction at the firm's expense. If customers experience this potential failure, they are to contact Sound Surgical Technologies at 1-888-677-8765 or 1-888-471-4777.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution -- AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, MA, MI, MO, NJ, NY, PA, and TX .