CLARAVUE Pre-wired Electrode - IP SET 5V RT ADULT Model: 32028781 50604-US
FDA Recall Z-0940-2022 · Posted None
Advertisement
Recalling FirmGraphic Controls Acquisition Corporation
Recall StatusTerminated
Product CodeDRX
Date Initiated2022-03-15
Date Terminated2024-07-12
Reason for Recall
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Firm Action
Graphic Controls (d/b/a Nissha Medical Technologies/Vermed/Biomedical Innovations) issued Urgent Field Correction letter via email on 3/4/22. Letter states reason for recall, health risk and action to take:
Section 5.1.5 of the standard provides that trunk cables and patient leadwires may be permanently marked to indicate they are not defibrillation proof. Immediately examine your inventory and complete the provided form Claravue IP Set Correction Notice Acknowledgement Form and return to [email protected] If you have product in inventory our team will arrange for it to be returned and exchanged for product labeled correctly once available. In addition, if you may have further distributed
this product, please identify your customer and notify them at once of this product correction. Your notification may be enhanced by including a copy of this correction letter. Please complete and return the attached form via e-mail to [email protected] as soon as possible.
If you have any questions about this notice, please contact:
Ms. Jessica Potter, Vice President Quality and Regulatory
Nissha Medical Technologies, 400 Exchange Street, Buffalo, NY 14204, USA
Phone: (716) 849-6367 E-Mail: [email protected]
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
Related FDA 510(k) Records