Terminated

DRAD-3000E: Radrex-i X-ray is a general purpose X-ray System. Labeling reads in part "TOSHIBA***DIGITAL RADIOGRAPHY SYSTEM RADREX-i DRAD-3000E Systems Data No. MSDXR0144EA***"

FDA Recall Z-0941-2012 · Posted 2012-02-02

Recall StatusTerminated
Product CodeMQB
Date Initiated2011-12-29
Date Terminated2014-03-24
Product Quantity39

Reason for Recall

It has been found that some of the starter cables used for x-ray tube high speed rotation are not sufficiently durable.

Firm Action

The recall notification titled "URGENT: MEDICAL DEVICE CORRECTION", was sent on 12/29/11 to customers via certified mail. The letter informed the customer of an issue with the starter cable used for high speed rotation on the Radrex X-Ray System. The letter also informs customers of the corrective measures: a replacement cable will be installed on your system to prevent this issue from occurring. An 800 number was provided for any questions regarding the letter.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Class II Recall - Nationwide Distribution (USA) - Including the states of VA, IL, CA, PA, TN, OH, LA, TX, OK, AR, AZ, MS, WV, TN, WI, NJ, MN, NY, IA, FL, and Washington D.C.