Large Combination Clamp/Large External Fixation System, Part Number 390.005 Product Usage: Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation.

FDA Recall Z-0941-2013 · Posted 2013-03-11

Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeKTT
Date Initiated2013-01-11
Date Terminated2015-08-31
Product Quantity136

Reason for Recall

The L1 hex nuts, used to tighten the Large Combination Clamp, were oversized and the 11 mm sockets and wrenches will not fit on the hex nuts.

Firm Action

Synthes sent an Urgent Notice: Medical Device Recall letter dated January 11, 2013 to all affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to examine their inventory for the affected product(s), remove them from inventory and return them to Synthes. For questions call 610-719-5450 or email [email protected]

Root Cause

Nonconforming Material/Component

Distribution Pattern

USA Nationwide Distribution including the states of AL, AR, AK, CA, FL, ID, KS, LA, MD, MA, MO, NJ, NY, OH, PA, TN, TX, and VA.