AccuSure Insulin Syringe, 1cc, 28 guage, 1/2'', packaged in boxes of 100, each containing 10 bundles of 10 syringes, disposable, NDC 0603-6996-21.
FDA Recall Z-0942-05 · Posted 2005-06-30
Advertisement
Recalling FirmQualitest Pharmaceuticals Inc
Recall StatusTerminated
Product CodeFMF
Date Initiated2005-06-14
Date Terminated2006-05-10
Product Quantityunknown
Reason for Recall
Labeling error. There is a possibility that 1cc syringes are labeled at 1/2cc on the plastic inner wrap holding bundles of 10 syringes.
Firm Action
The firm initiated the recall on June 14, 2005 with a press release and followed with a letter initially sent via fax and then by letter requesting return of the product and posting of a copy of the press release at the retail level.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.