ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1

FDA Recall Z-0942-2022 · Posted None

Recall StatusCompleted
Product CodeIYE
Date Initiated2022-03-21
Product Quantity14

Reason for Recall

Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location

Firm Action

On 3/21/22, Varian Medical Systems issued a "Urgent Medical Device Correction" notification to affected consignees. In addition to informing consignees about the recall, Varian asked consignees to take the following actions: Please retain a copy of this document along with your most current product labeling. Advise the appropriate personnel working in your radiotherapy department of the content of this letter. This document will be posted to the Varian Medical Systems customer support website: http://www.MyVarian.com. Please complete the attached Return Response form provided. Kindly return the completed form to [email protected].

Root Cause

Software Design Change

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of MI, KS, OH, VA, DC, FL, PA, OK, MN, AZ and the countries of South Korea, Netherlands, Poland, China.