MICROTEK(TM) MEDICAL BUR HOLE LATEX FREE PROBE COVER, featuring IsoSilk(R), 13cm x 244 cm, 5" x 96", REF PC3688N, QTY 100, Microtek Medical, Inc., Columbus, MS 39702.

FDA Recall Z-0943-2014 · Posted 2014-02-10

Recalling FirmEcolab Inc
Recall StatusTerminated
Product CodeKKX
Date Initiated2013-07-02
Date Terminated2014-05-14
Product Quantity1800 individual units

Reason for Recall

Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels.

Firm Action

The recall was conducted to the consumer/user level. The firm notified their direct consignees of the problem and the recall by written letter mailed on July 2, 2013. The notice requested the consignees to quarantine any recalled product on hand, and complete and return the response form by July 31, 2013. The consignees who further distributed the recalled product were asked to notify their consignees.

Root Cause

Process design

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Austria, Germany, Finland, Ireland, France, Spain, Great Britain, Belgium, Netherlands, Switzerland, Jordan, United Kingdom, Saudi Arabia, Hungary, Latvia, Malta, Norway, Poland, Portugal, and Canada.