PROFEMUR(R) R RASP HANDLE, REF PPW38078, 1 EACH, NON-STERILE, MicroPort Orthopedics, Arlington, TN Instruments support the implantation of implants. Broach handles mate to the broach to allow for preparation of the femoral canal for the stem component to be implanted. This
FDA Recall Z-0943-2017 · Posted 2017-01-03
Advertisement
Recalling FirmMicroPort Orthopedics Inc.
Recall StatusTerminated
Product CodeLZO
Date Initiated2016-11-17
Date Terminated2018-08-18
Product Quantity20 units
Reason for Recall
The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.
Firm Action
MicroPort Orthopedics sent an Voluntary Device Product Recall letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return the enclosed form even if they do not have any affected product to return. Return the completed form by fax to 1-901-451-6032 or by email to [email protected]. Attempt #1 for acknowledgment will be made by email. For questions regarding this recall call 901-867-4318.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution to Spain, China, and Italy
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.