Epicor LP Connecting Cable, CC-6-LP-CE, Catalog Number 12263, an accessory to the Epicor Cardiac Ablation Control System, manufactured by St.Jude Medical, Sunnyvale, CA Disposable connecting cable, accessory to the Epicor Cardiac Ablation Control System. The cable is used to connect the Epicor disposable tissue device (UltraCinch or UltraWand) to the Epicor Cardiac Ablation control System.

FDA Recall Z-0946-2009 · Posted 2009-01-30

Recalling FirmSt. Jude Medical Inc
Recall StatusTerminated
Product CodeOCL
Date Initiated2008-12-09
Date Terminated2010-12-28
Product Quantity158 units distributed.

Reason for Recall

Firm has received multiple reports of difficulties connecting the product and/or the presence of error code messages occurring with the use of the connecting cable and the Ablation Control System unit.

Firm Action

The firm issued notification by "Urgent Product Recall" letter dated 12/4/2008, to all consignees. The letter describes the issue, potential risk and states that the product should be returned. For additional information, contact St Jude Medical at 800-722-3423.

Root Cause

Other

Distribution Pattern

Worldwide - NY, IN, NC, UT, OR, CA, IL, MI, TX, TN, and FL and country of Belgium.