Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park, Illinois

FDA Recall Z-0948-2014 · Posted 2014-02-10

Recalling FirmEcolab Inc
Recall StatusTerminated
Product CodeKKX
Date Initiated2013-07-02
Date Terminated2014-05-14
Product Quantity6,000 individual units

Reason for Recall

Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels.

Firm Action

The recall was conducted to the consumer/user level. The firm notified their direct consignees of the problem and the recall by written letter mailed on July 2, 2013. The notice requested the consignees to quarantine any recalled product on hand, and complete and return the response form by July 31, 2013. The consignees who further distributed the recalled product were asked to notify their consignees.

Root Cause

Process design

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Austria, Germany, Finland, Ireland, France, Spain, Great Britain, Belgium, Netherlands, Switzerland, Jordan, United Kingdom, Saudi Arabia, Hungary, Latvia, Malta, Norway, Poland, Portugal, and Canada.