Merlin@home RF Remote Monitoring Transmitter, Model EX1150, with software versions 6.1M or 7.0. Usage:RF Remote Transmitter

FDA Recall Z-0949-2015 · Posted 2015-01-09

Recall StatusTerminated
Product CodeNVZ
Date Initiated2014-12-18
Date Terminated2016-03-30
Product Quantity257,456 units

Reason for Recall

Transmitter may initiate a software reset resulting in backup operation in some implanted St. Jude Medical Radio Frequency (RF) enabled Implantable Cardioverter Assura, Unify Assura, and Quadra Assura ICDs and Assurity and Allure Pacemakers.

Firm Action

The firm, St. Jude Medical, sent an "Important Medical Device Correction" Dear Doctor Letter dated 12/18/14 to customers to inform them that the Merlin@home transmitters may initiate a software reset resulting in backup operation in some implanted St. Jude Medical Radio Frequency (RF) enabled Implantable Cardioverter Defibrillators (ICDs) and Pacemakers. The letter informs the customers of the problems identified and the actions to be taken. Customers with questions or concerns are instructed to contact their local St. Jude Medical representative or St. Jude Medical's Technical Services Department at 1-800-722-3774.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: EU, Mideast, Canada, Australia/New Zealand, and Asia-Pacific (including Japan).