Evita 4 Continuous Ventilator

FDA Recall Z-0954-03 · Posted 2003-06-27

Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2003-05-07
Date Terminated2006-01-20
Product Quantity806 units

Reason for Recall

software problems - ventilation mode changes without user input, settings change without corresponding display change, false auto-set alarm.

Firm Action

The recalling firm issued recall letters to the accounts dated 5/7/03.

Root Cause

Other

Distribution Pattern

The product was shipped nationwide to distributor, hospitals and a physician. The product was also shipped to medical facilities and universities in Canada.