Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Covidien, LigaSure, Blunt Tip Laparoscopic Sealer/Divider, 5 mm-37 cm, REF LF1537, Sterile EO, Covidien, Valleylab, a division of Tyco Healthcare Group LP, 5920 Longbow Drive, Boulder, CO 80301 Intended use: LigaSure 5 mm devices are electrosurgical instruments designed to permanently seal and divide vessels, tissue bundles, and lymphatics.

FDA Recall Z-0954-2011 · Posted 2011-01-25

Recalling FirmCovidien LP
Recall StatusTerminated
Product CodeGEI
Date Initiated2010-12-08
Date Terminated2012-03-27
Product Quantity25,009 units

Reason for Recall

During surgical procedures, the device may energize when clamping without pushing the activation button, which could result in unintended energy delivery to tissue.

Firm Action

The firm, Covidien, sent a "URGENT: PRODUCT RECALL" letter dated December 6, 2010, to the customers. The letter described the product, problem and the action to be taken by the customers. The customers were instructed to remove product from inventory, immediately quarantine the product and return any remaining stock of the lots listed to: Covidien Energy-based devices, Attention: R82, Building 4/Receiving, 5920 Longbow Drive, Boulder, CO 80301. The customers were also instructed to complete and return the Customer Response Form immediately via mail (in the envelope provided) or fax to (303) 516-8307. To return any identified product for credit or replacement, contact Customer Service at 1-800-722-8772 Option 1. Should you have any medical questions concerning this recall, please contact your local sales representative or the Covidien Clinical information Hotline at (800) 255-8522 (option 1).

Root Cause

Device Design

Distribution Pattern

Worldwide distribution: USA and countries including: Australia, Canada, France, Germany, Japan, Puerto Rico, Singapore, and Uruguay.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑