Terminated

CONTACT RENEWAL 3 AVT model numbers M150 and M155; CONTACT RENEWAL 3 AVT HE model numbers M157 and M159; CONTACT RENEWAL 4 AVT model numbers M170 and M175; CONTACT RENEWAL 4 AVT HE model numbers M177 and M179. Cardiac Resynchronization Therapy Defibrillator. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA. CONTAK RENEWAK AVT models provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies.

FDA Recall Z-0956-05 · Posted 2005-06-30

Recall StatusTerminated
Product CodeLWP
Date Initiated2005-06-17
Date Terminated2008-03-02
Product Quantity2,204 devices (786 remain implanted worldwide)

Reason for Recall

The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy.

Firm Action

Firm issued a Dr. Doctor letter dated 06/17/05. Letter informs physicians of the issue, clinical implications, recommendations and action taken by the firm.

Root Cause

Other

Distribution Pattern

worldwide. To the US and OUS countries to include: Argentina, Australia, Austria, Belgium, Canada, Chile, Cyprus, Denmark, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Syrian Arab Republic, Turkey, US - Virgin Islands and United Kingdom.

Official Source

View FDA Recall →