MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.

FDA Recall Z-0956-2007 · Posted 2007-06-22

Recall StatusTerminated
Product CodeNPQ
Date Initiated2007-05-14
Date Terminated2009-12-17
Product Quantity4,201

Reason for Recall

False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these levels and could result in false positive results being reported.

Firm Action

Consignees were sent an Urgen Medical Device Correction letter dated 5/14/07 which instructs them to define and establish a lower detection limit, to use UNG to prevent carryover whenever possible, and to confirm low positive results through an independent experiment. A revised device correction letter dated 6/15/07 was subsequently sent to each consignee updating the reason for recall and associated risks.

Root Cause

Other

Distribution Pattern

Nationwide.