MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.
FDA Recall Z-0956-2007 · Posted 2007-06-22
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeNPQ
Date Initiated2007-05-14
Date Terminated2009-12-17
Product Quantity4,201
Reason for Recall
False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these levels and could result in false positive results being reported.
Firm Action
Consignees were sent an Urgen Medical Device Correction letter dated 5/14/07 which instructs them to define and establish a lower detection limit, to use UNG to prevent carryover whenever possible, and to confirm low positive results through an independent experiment. A revised device correction letter dated 6/15/07 was subsequently sent to each consignee updating the reason for recall and associated risks.
Root Cause
Other
Distribution Pattern
Nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.