Terminated

Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of a system indicated for the fixation of fractures, non-unions, arthrodeses, and osteotomies of bones appropriate for the size of the device.

FDA Recall Z-0960-2017 · Posted 2017-01-06

Recall StatusTerminated
Product CodeHWC
Date Initiated2016-12-08
Date Terminated2017-05-17
Product Quantity50 units

Reason for Recall

Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.

Firm Action

Trilliant Surgical Ltd. sent an Urgent Medical Device Recall letter dated December 8, 2016, to all affected consignees. The letter instructed the consignees to 1) identify any/all affected product; 2) document any/all identified product via a provided response form and return to Trilliant Surgical; 3) exchange product upon receipt of replacements; and 4) return any/all affected product as originally identified and reported on the return response form. Customers with questions were instructed to call 800-495-2919.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →