SRD S5 Mast Roller Pump System 150, S5 Component, Part No.: 50-80-00, Made in Germany, Distributed by: Sorin Group USA, Inc., Arvada, CO 80004. Used to perform, control and monitor extracorporeal blood circulation during an operation for up to six hours.

FDA Recall Z-0961-2011 · Posted 2011-01-26

Recalling FirmSorin Group USA, Inc.
Recall StatusTerminated
Product CodeDTQ
Date Initiated2010-11-12
Date Terminated2012-07-13
Product Quantity176 units

Reason for Recall

Touch screen may become unresponsive, inhibiting user input.

Firm Action

All affected US customers and distributors were notified by certified mail on Nov 12, 2010 via a Field Correction letter. They were told that they would be contacted by Sorin to arrange an appointment to inspect and replace affected touch screens on site. They were told that they could safely continue using their S5 systems in accordance with the Operator's Manual and previously supplied instructions until the inspection and necessary replacement had been completed. A customer response form was also included for completion and return to the firm. Customers can contact the firm if they have any questions.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution.