Heavy Duty PTFE Wire Guide, Guidewire, Global Product Number G14323 G14260 G14260

FDA Recall Z-0961-2020 · Posted None

Recalling FirmCook Inc.
Recall StatusTerminated
Product CodeOCY
Date Initiated2019-01-04
Date Terminated2020-05-29
Product Quantity146

Reason for Recall

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Firm Action

On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.

Root Cause

Process control

Distribution Pattern

Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.