DePuy CONFIDENCE SPINAL CEMENT SYSTEM 11C Plus KIT Catalog Number: 283910000. Delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures.

FDA Recall Z-0962-2013 · Posted 2013-03-18

Recalling FirmDePuy Spine, Inc.
Recall StatusTerminated
Product CodeNDN
Date Initiated2013-01-15
Date Terminated2014-06-30
Product Quantity7773

Reason for Recall

Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck

Firm Action

DePuy Synthes Spine issued notification letter dated 1/11/13 issued via UPS Next Day Air Saver on 17 January 2013. Accounts requested to review inventory, quarantine any products listed in this recall notice and to complete and return the enclosed Business Reply Form. Contact the firm at 508-977-6606 for questions relating to this notice.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Argentina, Armenia, Kazakhstan, Russia, Ukraine, Australia, Austria Belgium, Luxembourg , Brazil, Canada, China, Colombia, Cyprus, Jordan Lebanon, Saudi Arabia, UAE, Czech Republic, Finland ,France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Latvia, Malaysia Mexico, Netherlands, New Zealand, Norway, Sweden, Poland, Portugal Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Turkey, and UK.