Vitek 2 reagent cards: a) GN ID, REF 21341, b) GP ID, REF 21342, c) ANC ID, REF 21347, d) AST-GNOR, REF 22000, e) AST-GP67, REF 22226, f) AST-P612, REF 22359, g) AST-XN01, REF 410025, h) AST-N222, REF 413083, i) AST-N223, REF 413110, j) AST-N226, REF 413143, k) AST-N248, REF 413397, l) ASR-GN67, REF 413399, m) AST-GN69, REF 413400, n) AST-GN73, REF 413404, o) AST-GN79, REF 413436, p) AST-GN82, REF 413439, q) AST-N261, REF 413753, r) AST-XN06, REF 413944, s) AST-N275, REF 414313, t) AST-GP75, REF 415670, u) AST-N300, REF 416241, v) AST-N308, REF 413913, w) AST-N314, REF 417027, x) AST-N318, REF 417952, y) AST-P643, REF 418671, z) AST-YS08, REF 420739, aa) AST-N348, REF 420856, bb) AST-ST02, REF 420915, cc) AST-ST03, REF 421040, dd) AST-N355, REF 421351, ee) AST-P654, REF 421912, ff) AST-P655, REF 421913, gg) AST-N371, REF 422024, hh) AST-N372, REF 422241, ii) AST-N379, REF 423052,
FDA Recall Z-0962-2019 · Posted None
Advertisement
Recalling FirmOrganon Teknika Inc
Recall StatusTerminated
Product CodeLON
Date Initiated2019-01-11
Date Terminated2020-11-06
Product Quantity129,321 cartons
Reason for Recall
This notice has been initiated due to potential for compromised test card Top Seal Integrity which could:
- yield false resistance for antibiotics on the AST panel
- cause a false negative ESBL Test
- result in a false positive urea (URE) reaction on ID cards
Firm Action
The recall was issued to bioMrieux subsidiaries and distributors via email notification on Friday, January 11th, 2019.
Impacted customers were notified via correction notice (customer letter) issued on 01/19/2019 from each associated subsidiary/distributor onto local letterhead. The letter directed the user to perform a "tug test" to determine if the product is defective. Defective units were to be discarded and a credit issued.
Root Cause
Process control
Distribution Pattern
Worldwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.