TFE Coated Curved Newton LLT
FDA Recall Z-0963-2018 · Posted None
Advertisement
Recalling FirmCook Inc.
Recall StatusTerminated
Product CodeDQX
Date Initiated2017-10-13
Date Terminated2019-10-23
Product Quantity3,364
Reason for Recall
Label does not state that the product is heparin-coated.
Firm Action
On October 31, 2017 an URGENT MEDICAL DEVICE CORRECTION was issued to customers requesting the following, including a corrective action:
Step 1:
Determine if your product is
included in the affected lot
ranges by comparing your lot
number to the listing
provided.
" If your product is not
affected, complete the
Acknowledgement and
Receipt Form.
" If your product is affected,
proceed to Step 2.
Step 2:
Remove a Heparin Coated
label from the label sheet
provided and apply to the top
right corner of your affected
unit.
o If you need additional
labels, please contact
Stericycle at
855.215.4967.
If you would like assistance,
representatives are available to support
you. To request assistance, please
contact Stericycle at 855.215.4967
Root Cause
Labeling Change Control
Distribution Pattern
Nationally
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.