Puritan Bennett 980 Series Ventilator
FDA Recall Z-0966-2022 · Posted 2022-04-25
Advertisement
Recalling FirmMedtronic formerly Covidien
Recall StatusOpen, Classified
Product CodeCBK
Date Initiated2022-03-25
Product Quantity18,508 OUS; 18,412 US
Reason for Recall
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Firm Action
The firm notified consignees starting on March 25, 2022 via customer letter. Customers were informed of the issue and advised to institute workflow modifications to ensure visibility of the GUI visual alarm banner to ensure awareness of an active alarm condition in the absence of a functioning audible alarm and/or omni-directional LED visual alarm. If the audible alarm or omni-directional LED alarm fail, the clinical team should transfer the patient to another form of ventilation.
The firm will correct the issue in a future software update. For questions contact your Medtronic representative or Technical Service at 1-800-255-6774
The firm sent additional correction notices to consignees that received affected devices before the software correction was distributed.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide - Worldwide .Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.