LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

FDA Recall Z-0966-2026 · Posted 2025-12-19

Recalling FirmICU Medical, Inc.
Recall StatusOpen, Classified
Product CodePHC
Date Initiated2025-11-24
Product Quantity12 customers

Reason for Recall

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

Firm Action

On November 24, 2025, the firm notified customers via Urgent Medical Device Correction letters. Consignees were asked to take the following actions: 1. Ensure all users or potential users of LifeShield DLM 2.2.1 are immediately made aware of this notification and proposed mitigations. 2. Until a software update is released, limit any newly created or edited Concentration Limits to one decimal place of precision for limits less than 100. ICU Medical is developing a software update to correct this issue and will notify customers of the timing of the software update.

Root Cause

Software Design Change

Distribution Pattern

US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.