Medtronic U-CLIP Anastomotic Device Removal Tool, model REM1860FW. The U-Clip Anastomotic Device Removal Tool is designed to remove U-Clip Anastomotic Devices.

FDA Recall Z-0968-2007 · Posted 2007-06-27

Recall StatusTerminated
Product CodeFZP
Date Initiated2007-02-08
Date Terminated2008-02-24
Product Quantity187 (US 155, OUS 32)

Reason for Recall

Six lots of U-Clip Removal Tools, catalog REM1860FW were incorrectly assembled. The nose cone attachment to the body of the removal tool does not contain sufficient adhesive which can cause the nose to separate from the body of the device.

Firm Action

Customers to be notified via a letter on 2/8/07. The letter will explain the situation and requests that the impacted product be returned to Medtronic for disposal. The customer will be requested to complete a "Recall Certificate" acknowledging the receipt of this information.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.