Terminated

Halo One 5F Introducer Sheath 45 cm Product Code: HAL545 The Halo One Thin-Walled Guiding Sheath is designed to perform as both a guiding sheathand introducer sheath. The Halo One Thin-Walled Guiding Sheath is indicated for use inperipheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is NOT indicated for use in cerebrovascular nor the coronary vasculature

FDA Recall Z-0968-2017 · Posted 2017-01-18

Recall StatusTerminated
Product CodeDYB
Date Initiated2016-12-02
Date Terminated2017-04-05
Product Quantity101 units

Reason for Recall

Complaints of sheath separation, kinking, and/or tip damage during use.

Firm Action

Bard sent an Urgent Medical Device Recall Notification letter dated December 2, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to find attached instructions detailing the steps they need to take regarding the product. Customers with questions were instructed to contact their Bard representative. Customers were provided the following instructions: 1. Records show that your facility has purchased product codes affected by this voluntary recall. Do not use or further distribute any affected product. 2. Check all inventory locations within your institution for affected product code / lot number combination listed in the recall notice. If have further distributed any of the product code / lot numbers, immediately contact that location, advise them of the recall, forward instructions and have them return the affected product to BPV. 3. Remove any identified product from shelves. 4. If have used the affected product, complete and return the attached Recall and Effectiveness Check Form indicating no product will be returned. Once the product affected by this recall has been removed from inventory: 5. Fill out the Recall and Effectiveness Check Form. Be sure to state the quantities and lot numbers of each recalled product that intend to return. It is extremely important that this information is received . 6. Please call the firm's Recall Coordinator at 1-800-321-4254 Option #2 Ext 2501 (M-F 6am to 3pm MST). Once all information has been verified, the Recall Coordinator will issue you either a Return Authorization (XC) Number or Consignment Recall Number (XH) to facilitate the expedient return of the product. BPV will issue a credit for returned product. Email the completed Recall and Effectiveness Check Form to [email protected] or fax it to BPV at 1-800-994-6772. If cannot email or FAX the form, please call the BPV recall coordinator at 1-8

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution to NY, MO, NY, NY, FL, KS, MI, TX, NV, OH, ME, WA, LA, AZ, UT, NH. No foreign distribution.