DC Power Adapter for use with the LIFEPAK 12 defibrillator/monitor.
FDA Recall Z-0971-04 · Posted 2004-07-20
Advertisement
Recalling FirmMedtronic Physio Control Corp
Recall StatusTerminated
Product CodeLDD
Date Initiated2004-05-05
Date Terminated2005-01-06
Product Quantity1689
Reason for Recall
The DC Power Adapter may fail to charge LIFEPAK NiCd 2.4 AH batteries used with the LIFEPAK 12 defibrillator/monitor.
Firm Action
On 5/5/04 the firm issued a letter to consignees advising of the problem and that a service representative will contact them regarding replacement of the DC Power Adapter.
Root Cause
Other
Distribution Pattern
The firm distributed devices to medical facilities, fire departments (first responders).
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.