Fisher & Paykel Infinity Direct Nasal Mask, Part Numbers: HC481A & HC481SL. CPAP masks.
FDA Recall Z-0972-2007 · Posted 2007-06-27
Advertisement
Recalling FirmFisher & Paykel Healthcare Inc
Recall StatusTerminated
Product CodeBZD
Date Initiated2006-11-24
Date Terminated2012-03-26
Product Quantity815,239 total for all models, all lots
Reason for Recall
Plastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use.
In the event of a tab breaking off, it may enter the CPAP system air path and there is an extremely remote possibility that the broken tab may then enter the patient's nasal cavity or mouth. Masks manufactured since April 2006 do not feat
Firm Action
Fisher & Paykel Healthcare Inc. commenced sending out letters by mail to consignees the week commencing 4 December 2006 advising consignees to quarantine and destroy affected product. A faxback form was also be provided for consignees to advise Fisher & Paykel Healthcare of quarantined and destroyed stock.
The firm will supply consignees with replacement product as determined from completed faxback information.
The firm will collate responses from consignees and will follow up with a letter to consignees who do not respond within a month of initial mail-out. The firm will then make telephone contact with any consignees who do not respond within a month of the second mail-out.
Fisher & Paykel Healthcare Inc. estimated the completion of this voluntary recall by April 2007.
Root Cause
Other
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.