Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
FDA Recall Z-0972-2015 · Posted 2015-01-14
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2014-12-03
Date Terminated2015-04-06
Product Quantity50
Reason for Recall
Suboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in rare cases.
Firm Action
Siemens sent an important customer safety information notice, dated December 3, 2014, to all affected customers. The letter identified the product, problem, and action to be taken. The following hardware modifications will be implemented with the corrective action AX042/14/S:
1. A modified cable outlet to provide additional space for the cabling.
2. A modified plastic cover at the cable outlet providing optimal cable guidance in the area of the arc.
In general all systems will be checked for existing damage prior to the implementation of this action.
For questions regarding this recall call 610-219-6300.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution