Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.
FDA Recall Z-0973-2014 · Posted 2014-02-28
Advertisement
Recalling FirmMedtronic Navigation, Inc.
Recall StatusTerminated
Product CodeOXO
Date Initiated2013-07-25
Date Terminated2019-04-17
Product Quantity449 in US
Reason for Recall
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
Firm Action
Medtronic Navigation Inc. issued Notiifcation on 3/4/14 and included thee errata sheet to all affected customers updating the missing and incorrect information and ensure that labeling is now in compliance with 21 CFR Part 1020.30(h).
This will be followed by a software update that contains the same information on the errata sheet, which serves to correct the electronic version of the Instructions for Use.
The paper errata sheet will be sent via traceable means and the software upgrade will be performed by Medtronic Service either through remote access or by site visit of Medtronic service personnel.
For further questions please call (978) 698-6000.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
USA (nationwide) Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.