AU Clinical Chemistry Analyzers: Model Number: AU600 Clinical Chemistry Analyzer, AU640 Clinical Chemistry Analyzer, AU640(e) Clinical Chemistry Analyzer, AU680 Clinical Chemistry Analyzer. The AU Clinical Chemistry Analyzers are fully automated photometric analyzers intended for clinical laboratory use. Applications include colorimetric, turbidimetric, latex agglutination, and homogeneous enzyme immunoassay.

FDA Recall Z-0975-2012 · Posted 2012-02-07

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2010-07-12
Date Terminated2014-08-04
Product Quantity8225 units

Reason for Recall

The recall was initiated because Beckman has confirmed that the AU Clinical Chemistry Analyzers may experience cuvette overflow.

Firm Action

Beckman Coulter sent an "UPDATE TO PCA -14563: URGENT PRODUCT CORRECTION" letter dated April 27, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter is a communication which describes an alternative approach to detecting cuvette overflow. Additionally, a response form was enclosed for customers to complete and return to the firm. Contact the Customer Support Center at (800) 854-3633 or on the Beckman Coulter website at www.beckmancoulter.com/customersupport/support/default.asp for questions regarding this notification.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution-including USA (nationwide).