Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
FDA Recall Z-0975-2013 · Posted 2013-03-22
Recalling FirmOrthohelix Surgical Designs Inc
Recall StatusTerminated
Product CodeHTW
Date Initiated2012-10-30
Date Terminated2013-12-05
Product QuantityLot # 2399041, 97 pcs; Lot # 2399051, 143 pcs; Lot # 2399051R, 4 pcs & Lot # 2399061, 119 pcs
Reason for Recall
The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.
Firm Action
OrthoHelix sent a "FIELD REMOVAL NOTIFICATION" letter dated November 7, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 330-247-1444 for questions concerning this notice.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution including the states of CA, CO, IL, MI, MD, NC, NM, OH, OK, PA, SD, TX, UT, VA, WA and WI.