Horizon Medical Imaging, Picture Archive Communication System (PACS).

FDA Recall Z-0976-2014 · Posted 2014-02-12

Recall StatusTerminated
Product CodeLLZ
Date Initiated2013-12-23
Date Terminated2017-04-25
Product Quantity618 units

Reason for Recall

Images may be archived to an incorrect or invalid archive location resulting in incorrect images or no images being displayed for a study following retrieval from archive.

Firm Action

McKesson sent a field notice to customers with the potentially affected software versions on 12/23/2013.

Root Cause

Software design

Distribution Pattern

U.S. Nationwide (Puerto Rico), Australia, Canada, France, GU, IE, and The United Kingdom.