STA Neoplastine CI Plus (10), (cat. 0667). 12 x 10 ml. in vitro diagnostic for the determination of prothrombin time (PT).
FDA Recall Z-0979-2007 · Posted 2007-07-03
Advertisement
Recalling FirmDiagnostica Stago, Inc.
Recall StatusTerminated
Product CodeGJS
Date Initiated2007-03-30
Date Terminated2009-11-30
Product Quantity170 units
Reason for Recall
Based on the confirmed clotting times in the upper part of the assigned ranges of STA-Coag Control ABN control plasmas when used with STA-Neoplastine CI Plus, use of this lot may cause difficulty during the validation of the STA-Coag Control ABN.
Firm Action
Recall notification letters were sent on March 30, 2007 by first class mail. The letter advises customers that lot 061781 of STA-Neoplastine Cl Plus may cause difficulty during the vailidation process of STA-Coag Control ABN control plasmas, and asks that the lot number be discontinued.
Root Cause
Other
Distribution Pattern
Nationwide.