GEMINI TF Big Bore PET/CT, Computer Tomography System, Philips Healthcare System, Cleveland, OH

FDA Recall Z-0979-2014 · Posted 2014-02-12

Recall StatusTerminated
Product CodeKPS
Date Initiated2013-12-05
Date Terminated2015-08-25
Product Quantity47 Units

Reason for Recall

The TF Big Bore patient table experienced mechanical binding during horizontal table motion resulting in an automatic Emergency Stop (E-stop). An E-Stop will interrupt a current scan, resulting in an incomplete study.

Firm Action

ON 12/5/2013 the firm sent Medical Device Correction Letters to their customers.

Root Cause

Software design

Distribution Pattern

Product was shipped to the following states: AR, CA, CO, FL, IL, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, PA, TX, VA, VT & WI. Product was also shipped to the following countries: Belgium, Canada, China, Denmark, France, Germany, Italy, Japan, Netherlands, Saudi Arabia & Venezuela.