Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364. For transseptal left heart access in both diagnostic and interventional procedures.

FDA Recall Z-0979-2019 · Posted 2019-03-13

Recalling FirmCook Inc.
Recall StatusTerminated
Product CodeDRC
Date Initiated2019-02-01
Date Terminated2020-06-30
Product Quantity100

Reason for Recall

The products were manufactured without a back bevel on the needle tip. This could cause damage to the inside of the introducer during needle insertion.

Firm Action

Urgent Medical Device Recall notification letters were sent to customers on 2/1/19. Action to take 1. Examine your inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of these products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. NOTE: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time) or by email to [email protected].

Root Cause

Process control

Distribution Pattern

US nationwide and Canada. US distribution to the following states: AZ, FL, IL, IN, KS, MA, MI, MO, PA, SC, TN, and WA. The products were distributed to the following foreign countries: Canada.