DePuy DNP Anatomic Fracture Repair System, Locking Screw DNP Anatomic, Depuy, Warsaw, IN; REF DNPALS. The device is used in the treatment of fractures and osteotomies of the distal radius.
FDA Recall Z-0981-2009 · Posted 2009-02-11
Advertisement
Recalling FirmDepuy Orthopaedics, Inc.
Recall StatusTerminated
Product CodeHRS
Date Initiated2008-12-23
Date Terminated2010-10-18
Product QuantityApprox. 3,250 of all products
Reason for Recall
The DNP nail plate may not attach to the alignment guide (jig) instrument, making it difficult to insert the nail plate into position or to drill needed holes accurately.
Firm Action
The firm's sales force was instructed via "Urgent Device Correction" letter issued December 29, 2008 to inspect all of these instruments in their possession, and those instruments at their hospital customers, and to leave a copy of the letter with each hospital customer who has the instruments on site.
The letter describes the issue and corrective actions for customers. Please direct questions about this recall to Depuy Orthopaedics, Inc. at 888-783-7156.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.