Open, Classified

ORAcollect Dx, Catalog Number/SKU: OCD-100, OCD-100A. Material numbers: OCD-100, OCD-100.008, OCD-100.012, OCD-100.014+ACP078, OCD-100.014+ACP078, OCD-100.032 and OCD-100A, OCD-100A.022, OCD-100A.023, OCD-100A.024, OCD-100A.025

FDA Recall Z-0981-2022 · Posted None

Recalling FirmDNA Genotek Inc.
Recall StatusOpen, Classified
Product CodeOYJ
Date Initiated2022-02-18
Product Quantity2,375,514 (US: 2,148,222 OUS: 227,292)

Reason for Recall

Evaporation of stabilizing liquid reducing shelf life.

Firm Action

Notification emailed to customers 2/18/2022. The shelf-life "Collect sample by/Use by" date found on the collection device tube label has been reduced by 12 months. Contact firm if unable to recover the minimum input volume required for sample extraction. Submit online acknowledgement form and contact your DNA Genotek Account manager or Technical Services with questions: [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WI and the countries of Australia Belgium, Brazil, Canada, Chile, China, Croatia Denmark, Finland, Germany, Great Britain, Italy, Japan, Netherlands, South Africa, and Thailand.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →