Terminated

Intera MRI

FDA Recall Z-0982-2007 · Posted 2007-06-27

Recall StatusTerminated
Product CodeLNH
Date Initiated2007-04-30
Date Terminated2011-09-21
Product Quantity662 units

Reason for Recall

Potential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.

Firm Action

On 4/30/07 Philips sent an Urgent Device Correction letter dated April 25, 2007 via certified mail to customers. The letter advises customers of the issue and that that a representative will visit the sites, inspect and correct the problem if indicated.

Root Cause

Other

Distribution Pattern

Units were distributed throughout the US to 1,078 hospitals and medical centers.