Open, Classified

Box as labeled: Regard, Carter-Thomason CloseSure System XL, REF 733776, Quantity 1 Box of 3 Systems. Kit as labeled: Regard, Carter-Thomason CloseSure System, REF 733776, Contents: One (1) Carter-Thomason Suture Passer, One (1) 5 mm Pilot Guide, One (1) 10/12 mm Pilot Guide.

FDA Recall Z-0985-2024 · Posted 2024-02-05

Recalling FirmCooperSurgical, Inc.
Recall StatusOpen, Classified
Product CodeGCJ
Date Initiated2023-08-29
Product Quantity127 boxes, 3 kits per box (381 kits total)

Reason for Recall

Incorrect component descriptions within the "Contents" section on the Tyvek lid for individual kit trays.

Firm Action

An Urgent Medical Device Field Safety Notice dated August 29, 2023 was sent to the one distributor receiving shipments of the affected product lot. Customers are advised to complete the acknowledgement form, Affected product is to be returned to CooperSugical for credit. Contact CooperSurgical at +1-203-601-5200 extension 3300 during normal operating hours of 09:00 - 17:00 M-F EST, email [email protected].

Root Cause

Labeling Change Control

Distribution Pattern

US Nationwide distribution in the state of MO.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →