Open, Classified

CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

FDA Recall Z-0988-2025 · Posted 2025-01-17

Recalling FirmSt. Jude Medical
Recall StatusOpen, Classified
Product CodeMOM
Date Initiated2024-10-05
Product Quantity605 units

Reason for Recall

As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.

Firm Action

On 11/14-18/24, the firm sent an updated Notification informing customer that Abbott is recommending replacement of the affected Patient Electronics System (PES) that are resisting correction of their configuration file. On 12/13-16/2024, the firm sent an updated Notification for the Hospital Electronic Systems (HES) regarding inaccurate Pulmonary Artery (PA) reading. On October 15, 2024, Abbott initiated a database correction to return affected PES to their intended configuration files. After continued monitoring, Abbott has identified a subset of units that is resisting correction and is retaining their incorrect configuration file. To ensure impacted PES units return to transmission of accurate data, Abbott recommends that clinicians take the following actions: 1 Review the attached appendix which includes a complete listing of impacted patients PES systems. -Patients with a date listed have already successfully connected and resumed sending accurate data. This date corresponds to the return of accurate PA pressures being sent by the PES. From this date forward, the data may be used for patient management. -Patients without a date listed have not yet resumed sending accurate data. Please encourage patients to send daily data transmissions (PES readings) to ensure the correction can be initiated. 2 PA pressures taken between 9/30/2024 and the listed date on the attachment may still not be accurate even after the database correction and it is recommended not to use for patient management. For HES units, review the appendix to see patient sensors that have been impacted by the affected HES between September 30, 2024 and the HES correction date, and take action as recommended below: 1 If the HES was used to take a follow-up pressure reading it is recommended to ignore that reading for clinical decision making or patient management. 2 If the affected HES was used for implant or recalibration, please contact Abbott Rep or Remote Care Technical Support at 1-844

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NV, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and the countries of Australia, and Canada.

Official Source

View FDA Recall →