GEHC OEC 9900 ESP-15 Elite C-Arm fluoroscopic X-ray system, GE Healthcare, Salt Lake City, UT.

FDA Recall Z-0989-2007 · Posted 2007-07-12

Recall StatusTerminated
Product CodeJAA
Date Initiated2007-07-06
Date Terminated2013-02-01
Product Quantity22 units

Reason for Recall

The Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.

Firm Action

Consignees were notified by certified letter on 07/06/2007. They were told that a service rep would contact them to provide a label and an addendum to their operator's manual. A permanent software solution is under development and will be provided when it becomes available.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, France, Singapore, Sweden, and Switzerland.