Terminated

i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448

FDA Recall Z-0989-2017 · Posted 2017-01-10

Recall StatusTerminated
Product CodeNEP
Date Initiated2016-11-23
Date Terminated2017-03-15
Product Quantity121 systems

Reason for Recall

The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.

Firm Action

Omincell mailed an Urgent Medical Device Field Correction Notification letter to affected customers on 11/21/2016. The letter provided information of the event and instructions on how to continue operating the device to prevent risks. The issue will be corrected with a software update.

Root Cause

Software design

Distribution Pattern

Shipped throughout United States.