i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448
FDA Recall Z-0989-2017 · Posted 2017-01-10
Advertisement
Recalling FirmHEALTH ROBOTICS S.R.L.
Recall StatusTerminated
Product CodeNEP
Date Initiated2016-11-23
Date Terminated2017-03-15
Product Quantity121 systems
Reason for Recall
The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.
Firm Action
Omincell mailed an Urgent Medical Device Field Correction Notification letter to affected customers on 11/21/2016. The letter provided information of the event and instructions on how to continue operating the device to prevent risks. The issue will be corrected with a software update.
Root Cause
Software design
Distribution Pattern
Shipped throughout United States.