Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q

FDA Recall Z-0989-2022 · Posted None

Recall StatusOpen, Classified
Product CodeLWS
Date Initiated2022-03-10
Product Quantity2

Reason for Recall

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Firm Action

Abbott issued a Safety Notification For a subset of Gallant ICDS and CRT-D, Models CDVRA500Q, CDDRA500Q and CDHFA500Q on March 10, 2022. Notification was hand delivered in US. The notification includes the Patient Management Recommendations: -As the devices are unable to be followed using remote monitoring via Bluetooth connectivity, follow-ups should be conducted in-clinic using inductive telemetry. - Abbott has developed a process to attempt recovery of Bluetooth communication via a device firmware upgrade. If remote monitoring via Bluetooth communication is desired, please work with your local Abbott representative to schedule a firmware upgrade. Please complete and return the included Acknowledgement Form to [email protected] and maintain a record of this notice to ensure effectiveness of the communication. For any question, please contact Abbott Technical Support at 1-800-722-3774 (U.S.).

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.

Official Source

View FDA Recall →